ATOLL-1M

Purpose:
  • registration of gamma-ray spectra and identification of radionuclides;
  • measuring radionuclide activity in building materials, mineral raw materials, waste products, food stuff and other materials in the Vessel Marinelli geometry.
The gamma-ray spectrometer "ATOLL-1M" can operate both in laboratory conditions (with low- background protection) and in field conditions. The instrument has an effective system of thermal compensation.

Delivery set:
The gamma-ray spectrometer "ATOLL-1M", notebook; specialized software; USB-cable; set of operational documentation, certificate about metrological documentation.

Specifications
Range of registered of gamma-quantum energyfrom 0.05 to 3.0 MeV
Relative energy resolution of Gamma-ray  line Cs137Not more than 7,5 %
Integral nonlinearity, %Not more than ±1
Maximum permissible major relative measurement error of samples activity in 1 litre Marinelli vessel at 1 hour exposure time , %Not more than ±30
Minimum detectable activity of Cs137 at 1 hour exposure time, Bq/kgNot more than 2 ( in 10-fold shielded chamber )
Detector materialNal(TI) diam.63x63
Version of detector unitSealed
Maximum statistical charge , c-1Not less than 104
Number of channels of the amplitude analyser1024
Channel capacity1000000
Reducing integral background count of detector block in the low background protection10 times
The gamma-ray spectrometer "ATOLL-1M" is powered from computer by means of USB cable, consumption currentNot more than 100 mA
Mode of operationRound-the-clock
Average lifetime of spectrometer6 years

Operation conditions:
ambient temperatureFrom minus 20°C up to 35°C
relative humidityUp to 95 %
Warranty period18 months

New algorithm of thermal compensation realized in spectrometer enable to carry out measurements in conditions of changeable temperature without calibration.скачать dle 12.0